Input PowerAC380V ±10% / 50Hz   
ResolutionAD:16bit;DA:16bit
Input Resistance≥1MΩ
Input Power7089W
Channel FeaturesConstant Current Source And Constant Voltage Source With Independent Pairs Of Closed-Loop Structure
Channel Control ModeIndependent Control
VoltageConstant Voltage Control Range0.025V~5V
Lowest Discharge Voltage2.5V
Accuracy± 0.1% Of FS
Stability± 0.1% Of FS
CurrentPer Channel Output Current Range0.5A-100A
Cut-Off Current Of CV0.2A
Accuracy± 0.1% Of FS
Stability± 0.1% Of FS
PowerPer Channel Output Power500W
Stability± 0.2% Of FS
TimeCurrent Response TimeThe Current Gets To 10%-90% Of FS Hardware Response Time Is 20ms
Step Time RangeSingle Step Time Range≤365*24 H,Time Form Support:00:00:00.000(H、Min、S、Ms)
Data RecordingRecord ConditionMin Time△T:100ms
Min Voltage△U:10mV
Min Current△I:200mA
Record Frequency10Hz
ChargeCharge ModeConstant Current Charge、Constant Voltage Charge、Constant Current And Constant Voltage Charge、Constant Power Charge
End ConditonVoltage、Current、Relative Time、Capacity、-△V
DischargeDischarge ModeConstant Current Discharge、Constant Power Discharge、Constant Resistance Discharge
End ConditionVoltage、Current、Relative Time、Capacity
PulseChargeCCC,CPC
DischargeCCD,CPD
Minimum Pulse Width500ms
Pulse32 Pulse Setting In One Step
Automated SwitchAutomated Switch From Charge To Discharge For Each Pulse
End  ConditionVoltage, Test Time
L  Setting Protection Condition,Setting Parameter:Low Voltage Limit、Upper
DCIRSupport  DCIR  Testing
ParallelMax 4 Channels Parallel, Don’t Support  Pulse Test After Parallel
CycleLoop Measure Range1~65535 Times
Max Steps Per Loop254
Nested LoopNested Loop Function,Max Support 3 Layers
ProtectionSoftware ProtectionL  Power-Down Data Protection
L  Off-Line Testing Function
L  Setting Protection Condition,Setting Parameter:Low Voltage Limit、Upper Voltage Limit、Low Current Limit、Upper Current Limit、Delay Time
Hardware ProtectionAdditional Anti-Reverse Protection Model
Voltage And Current Testing Sample4-Wire Connecting
Noise<85dB
Data OutputEXCEL2003、2010、TXT
Communication InterfaceEthernet Port
Per Unit Main Channels No.8