Description
Input Power | AC380V ±10% / 50Hz | |
Resolution | AD:16bit;DA:16bit | |
Input Resistance | ≥1MΩ | |
Input Power | 7089W | |
Channel Features | Constant Current Source And Constant Voltage Source With Independent Pairs Of Closed-Loop Structure | |
Channel Control Mode | Independent Control | |
Voltage | Constant Voltage Control Range | 0.025V~5V |
Lowest Discharge Voltage | 2.5V | |
Accuracy | ± 0.1% Of FS | |
Stability | ± 0.1% Of FS | |
Current | Per Channel Output Current Range | 0.5A-100A |
Cut-Off Current Of CV | 0.2A | |
Accuracy | ± 0.1% Of FS | |
Stability | ± 0.1% Of FS | |
Power | Per Channel Output Power | 500W |
Stability | ± 0.2% Of FS | |
Time | Current Response Time | The Current Gets To 10%-90% Of FS Hardware Response Time Is 20ms |
Step Time Range | Single Step Time Range≤365*24 H,Time Form Support:00:00:00.000(H、Min、S、Ms) | |
Data Recording | Record Condition | Min Time△T:100ms |
Min Voltage△U:10mV | ||
Min Current△I:200mA | ||
Record Frequency | 10Hz | |
Charge | Charge Mode | Constant Current Charge、Constant Voltage Charge、Constant Current And Constant Voltage Charge、Constant Power Charge |
End Conditon | Voltage、Current、Relative Time、Capacity、-△V | |
Discharge | Discharge Mode | Constant Current Discharge、Constant Power Discharge、Constant Resistance Discharge |
End Condition | Voltage、Current、Relative Time、Capacity | |
Pulse | Charge | CCC,CPC |
Discharge | CCD,CPD | |
Minimum Pulse Width | 500ms | |
Pulse | 32 Pulse Setting In One Step | |
Automated Switch | Automated Switch From Charge To Discharge For Each Pulse | |
End Condition | Voltage, Test Time | |
L Setting Protection Condition,Setting Parameter:Low Voltage Limit、Upper | ||
DCIR | Support DCIR Testing | |
Parallel | Max 4 Channels Parallel, Don’t Support Pulse Test After Parallel | |
Cycle | Loop Measure Range | 1~65535 Times |
Max Steps Per Loop | 254 | |
Nested Loop | Nested Loop Function,Max Support 3 Layers | |
Protection | Software Protection | L Power-Down Data Protection |
L Off-Line Testing Function | ||
L Setting Protection Condition,Setting Parameter:Low Voltage Limit、Upper Voltage Limit、Low Current Limit、Upper Current Limit、Delay Time | ||
Hardware Protection | Additional Anti-Reverse Protection Model | |
Voltage And Current Testing Sample | 4-Wire Connecting | |
Noise | <85dB | |
Data Output | EXCEL2003、2010、TXT | |
Communication Interface | Ethernet Port | |
Per Unit Main Channels No. | 8 |
Click one of our contacts below to chat on WhatsApp